Bticino room expander module, connection of up to 4 additional patient pushbuttons or bathroom pulls, for advanced hospital systems, with the following features:
Expansion: Up to 4 additional patient pushbuttons or bathroom tie-rods in addition to the basic one
Compatible with: CMTE1100 and CMTE1200 room terminals
Power supply: Derived from terminals, no external source required
Mounting: 6-module DIN rail
Low power consumption: 50 mA (stand-by), 65 mA (max)
Degree of protection: IP20, for installation in protected environment
(modificale nel modulo Rassicurazioni cliente)
(modificale nel modulo Rassicurazioni cliente)
(modificale nel modulo Rassicurazioni cliente)
The room expander module is a device designed to expand the functionality of room terminals in nurse call and patient management installations, allowing up to 4 patient pushbuttons or bathroom pulls to be connected in addition to the one already integrated. This makes multi-bed or multi-service management possible, ideal for shared hospital rooms or assisted bathrooms.
The module is fully compatible with the CMTE1100 and CMTE1200 room terminals, from which it also receives the power necessary for its operation. Thanks to this integration, it is easily installed without the need for additional power supplies, simplifying system design.
The device has a footprint of 6 DIN modules, and can be mounted directly inside switchboards dedicated to the call system. Built-in terminals facilitate wiring with additional pushbutton panels, while dip switches allow logical configuration of the number of beds or bathroom rods connected.
Key features include:
Signal LED, which flashes during module activation
Reset button, useful for resetting factory parameters
Dedicated terminals for connecting additional pushbuttons or bathroom tie-rods
Dip switches for identification of beds or bathroom stations (up to 4 devices)
The room expander module is designed for environments such as:
Hospital wards with multi-patient rooms
Assisted living facilities (RSAs)
Public and private health care facilities
Inpatient areas with multiple bathroom rows
Rooms with special patient care needs
| Parameter | Value |
|---|---|
| Reference Standards | EN60598-1, EN60601-1-2, EN11197, EN12464-1 |
| Supply voltage | Derived from BUS/SCS of CMTE1100 and CMTE1200 terminals (no external source required) |
| Power consumption | 50 mA (standby) / 65 mA (max) |
| Degree of protection | IP20 |
| Operating temperature | +5/+40 degrees |
| Installation | On DIN rail - 6 modules |
| Available interfaces | Terminals for patient pushbuttons 1-2 or bathroom tie-rods 2-3 Terminals for pushbuttons 3-4 or bathroom tie-rods 4-5 |
| Configuration | Via dip switches (bed/dial numbering) |
| Additional functions | Signal LEDs, reset button |
| Expansion | Up to 4 patient pushbuttons or bathroom pulls in addition to the basic one |
| Compatibility | CMTE1100 and CMTE1200 room terminals |